Abstract.Recommended best practices in monitoring of product status during pharmaceutical freeze drying are presented, focusing on methods that apply to both laboratory and production scale. With respect to product temperature measurement, sources of uncertainty associated with any type of measurement probe are discussed, as well as important differences between the two most common types of temperature-measuring instruments-thermocouples and resistance temperature detectors (RTD). Two types of pressure transducers are discussed-thermal conductivity-type gauges and capacitance manometers, with the Pirani gauge being the thermal conductivity-type gauge of choice. It is recommended that both types of pressure gauge be used on both the product chamber and the condenser for freeze dryers with an external condenser, and the reasoning for this recommendation is discussed. Developing technology for process monitoring worthy of further investigation is also briefly reviewed, including wireless product temperature monitoring, tunable diode laser absorption spectroscopy at manufacturing scale, heat flux measurement, and mass spectrometry as process monitoring tools.KEY WORDS: heat flux measurement; mass spectrometry; pressure measurement; process analytical technology; temperature measurement; tunable diode laser absorption spectroscopy.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.