The AutoMate Express™ Forensic DNA Extraction System was developed for automatic isolation of DNA from a variety of forensic biological samples. The performance of the system was investigated using a wide range of biological samples. Depending on the sample type, either PrepFiler™ lysis buffer or PrepFiler BTA™ lysis buffer was used to lyse the samples. After lysis and removal of the substrate using LySep™ column, the lysate in the sample tubes were loaded onto AutoMate Express™ instrument and DNA was extracted using one of the two instrument extraction protocols. Our study showed that DNA was recovered from as little as 0.025 μL of blood. DNA extracted from casework-type samples was free of detectable PCR inhibitors and the short tandem repeat profiles were complete, conclusive, and devoid of any PCR artifacts. The system also showed consistent performance from day-to-day operation.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.