INTRODUCTION: With rising age, functional deficit and frequent falls may lead to long-term care admission. Mobility assessment tests can detect fall risk and may induce interventions that prevent a fall.OBJECTIVES: To assess mobility of older persons using real time data and to compare these data with the mobility assessment of physiotherapists.METHODS: 20 older people aged 74±5 (mean ± SD) were monitored over 10 months to investigate the performance of an automated mobility tracker. Physiotherapists performed periodic mobility assessments. Annotated 3d recordings served as ground truth data.RESULTS: High correlation (r=0.684) of annotated and tracked gait speed was found. The mean absolute error is 0.16 m/s.CONCLUSION: 3D mobility trackers can be used to collect long-term mobility data. Since changes in mobility might indicate functional decline, long-term tracking allows to react to changes in mobility. Such a technology may have essential medical and social value.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.