We automated the Jendrassik-Grof method for measurement of bilirubin with the Greiner Selective Analyzer GSA II D. Comparison with the manual Jendrassik-Grof method showed a highly significant correlation coefficient of 0.994 for total and 0.980 for "direct" bilirubin measurement. The automated method showed good precision and was linear to 667 mu mol/L, the highest concentration tested. Addition of hemolysate caused a greater effect on the measurement of direct bilirubin than on total bilirubin. The inhibitory effect on measurement of total bilirubin, though small, was statistically significant at hemoglobin concentrations greater than 2 g/L. Comparison with the method involving diazotized 2,4-dichloroaniline showed the Jendrassik-Grof method to be relatively less affected by hemolysis and turbidity in the measurement of total bilirubin.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.