Background: Conventional manual compression relies on the surgeon's subjective sensations, so excessive compression can cause tissue injury to the stapling line of the intestinal anastomosis.Methods: Automatic compression monitoring and compression control system was developed for circular stapler. The tissue injury related compression variables were evaluated and accommodated by compression control device. The compression injury-reducing performance was verified on collagen sheets of in vitro experiments.Results: Excessive pressure and tissue deformation were associated with compression-induced tissue damages. The safe pressure range was very narrow in weaker tissue than normal collagen. The automatic system performed proper compression within a safe pressure range without tissue injury.Conclusions: Manual compression of circular stapler could cause tissue injuries by excessive pressure and tissue deformation. Our automatic compression system is designed to control peak pressure to prevent the compressive tissue injury.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.