The method describes concentration of mercury in samples of drinking water from water mains with Hg concentration above 1.0 . 10-7 g l-1 (5.0 . 10-10mol l-1) by means of extraction chromatography on a column packed with a carrier Synachrom E-5 wetted with saturated solution of bis(diethyldithiocarbamate)copper(II) complex in 1 : 1 mixture 1,2-dichlorobenzene + cyclohexane. After elution with 3.0 mol l-1 HCl the mercury concentration in the eluate is determined by atomic absorption spectrometry - cold vapour method. Average yield of the mercury separation from the model solutions was 95.8%.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.