Closed cycle He 3 -He 4 dilution cryostats became the platform of choice in quantum sciences in the era of helium shortage. However, in many experiments, the mechanical vibrations induced by the pulsed cryocoolers present a significant drawback reflected both in electronic and mechanical noises. Here, we present a hybrid dilution cryostat platform; we have automated a commercial closed-cycle system to operate on a cryocooler or on a liquid helium 'battery'. We implemented a scanning SQUID microscope in the hybrid dilution refrigerator. In this work we show the design of the hybrid setup and how its operation eliminates vibration artefacts in magnetic imaging.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.