Techniques are described for the design of shielded gradient coils for superconducting MRI systems. These design methods are suited for constructing the most efficient gradient coil that meets a specified homogeneity requirement. Tradeoffs in coil design of efficiency with coil size and gap size are discussed. Residual eddy currents from coils constructed with a finite number of wires are calculated and give guidelines for the construction of efficient, whole-body gradient coils.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.