A new sensing technology platform integrating magnetic permeability detection and a two-site heterogeneous immunoassay in a one-step analysis is described. As a platform model, measurements of C-reactive protein (CRP), a cardiac and inflammation marker, were performed in a rapid (11.5 min) high-sensitivity (hs) procedure with a low detection limit (0.2 mg/L) and accuracy (CV = 11%). The two-site heterogeneous immunoassay was performed in 1.2-mL disposable reagents vials containing solid phase (polyclonal anti-CRP conjugated silica microparticles), labeling agent (monoclonal anti-CRP conjugated superparamagnetic nanoparticles), and reaction buffer. Whole blood (20 muL) was assayed by introducing the sample into a reagent vial using a glass capillary and mixing its contents by hand for 30 s. After a 11-min sedimentation step, the vial was placed into the coil of the magnetic permeability detector, which measured the enrichment of superparamagnetic nanoparticles in the solid-phase sediment. Magnetic permeability detection and quantification is based on the principle that when paramagnetic materials are placed inside a coil, the inductance of the coil is influenced. Screening of CRP on whole blood patient samples showed good correlation with central hospital measurements for hsCRP (y = 1.018x - 0.021, R(2) = 0.980, n = 103) and normal range CRP (y = 1.02x + 2.53, R(2)= 0.991, n = 33) analyses. The mean differences of the two methods according to the Bland and Altman plots were -0.03 +/- 1.12 mg/L for hsCRP analysis and -3.4 +/- 8.64 mg/L for normal range CRP analysis.
A rapid (5.5 min) one-step whole blood C-reactive protein (CRP) magnetic permeability immunoassay utilizing monoclonal antibody conjugated dextran iron oxide nanoparticles (70 nm) as superparamagnetic labels and mixed fractions (1:1 ratio of 15-40 and 60 microm) of polyclonal anti-CRP conjugated silica microparticles for enhanced sedimentation is described. In this one-step assay procedure, a whole blood sample (4 microl) is applied to an assay glass vial, containing both antibody conjugates, and mixed for 30 s. The target analyte, CRP, forms a sandwich complex between the conjugated nanoparticles and microparticles, and, subsequently, the complex sediments under normal gravitation within 5 min to the bottom of the vial. The magnetic permeability increase of the sediment due to the presence of the complexed superparamagnetic nanoparticles is determined using an inductance-based transducer. Assayed patient whole blood samples were compared with the Abbott Diagnostics Architect reference method. A strong linear correlation was observed for the CRP concentration range 0-260 mg/l in whole blood (y=1.001x+0.42, R2=0.982, n=50). The CRP assay presented showed a limit of detection of 3 mg/l and a total imprecision (coefficient of variation) of 10.5%. On the basis of our observations, we propose a rapid, one-step, CRP assay for near-patient testing.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.