A pH-sensitive detector for flow-through potentiometry based on a graphite/quinhydrone/silicone composite electrode was applied to determine calcium and magnesium in aqueous solutions. The determinations are performed by FIA titrations with EDTA as the carrier solution. The method was applied to the analysis of mineral water, drinking water, and river water. The results were compared with those obtained by batch complexometric titration. The proposed new method is suitable for the simple, rapid and automated determination of water hardness in small sample volumes of aqueous solutions. The detection limit for both ions is about 1 10 À5 mol L À1 , the standard deviation is less than 2%. The sampling frequency is 60 determinations per hour.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.