This study aimed to evaluate the mechanical properties of contemporary heat-treated nickel-titanium (NiTi) rotary instruments used in the single-length technique [ProTaper Next (PTN), HyFlex EDM (EDM), and JIZAI (JZ)]. Bending loads, cyclic fatigue resistance, torque/force values and canal-centering ratios were evaluated for the three instruments and a non-heat-treated experimental NiTi instrument with the same geometry as JZ (nJZ). EDM and JZ exhibited significantly lower bending load and more cycles to failure compared with nJZ and PTN (p<0.05). PTN and JZ exhibited significantly better centering ability than nJZ and EDM (p<0.05). JZ and nJZ generated significantly smaller upward force and maximum torque than PTN and EDM (p<0.05). Under the present experimental condition, JZ exhibited flexibility and cyclic fatigue resistance comparable to EDM, better maintained the canal curvature than the other instruments, and generated smaller torque and screw-in force than PTN and EDM.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.