Objectives: To quantify and validate the effect of pixel size on a digital intraoral radiographic system according to International Electrotechnical Commission standards through physical and visual evaluations. Methods: The digital intraoral radiographic system used was the photostimulable phosphor imaging plate and scanner system. The system had three image capture modes: high-speed (HS), high-resolution (HR), and super high-resolution (SHR) with different pixels. The physical characteristics of the system were evaluated using presampled modulation transfer function (MTF) and the normalized noise power spectrum (NNPS). An aluminum (Al) step phantom with different depths of holes was used to acquire images under various exposure conditions. The average number of perceptible holes from all steps was plotted against each exposure dose. The results were compared to analyze the effects of pixel size on image quality of intraoral radiographs. Results: The MTF was slightly higher with SHR than with HR and HS. The NNPS with SHR showed about a 40% decrease in magnitude compared to HS. The total number of perceptible holes in the Al step phantom was higher with SHR than with HS and HR in all exposure conditions. Conclusions: The MTF and NNPS obtained with different pixel size could be quantified by physical evaluation, and the differences were visually validated with Al step phantom. The SHR mode has the potential to decrease the radiation dose without compromising the image quality.
Purpose:
To develop a method for improving sharpness of images reproduced on liquid‐crystal displays (LCDs) by compensating for the degradation of modulation transfer function (MTF) of the LCD.
Methods:
The inherent MTF of a color LCD (display MTF) was measured using a commercially available color digital camera. The frequency responses necessary to compensate for the resolution property of the LCD were calculated from the inverses of the display MTFs in both the horizontal and vertical directions. In addition, the inverses of the display MTFs were combined with the response of the human eye. The finite impulse response (FIR) filters were computed by taking the inverse Fourier transform of the frequency responses, and the effects of the FIR filtering on both the resolution and noise properties of the displayed images were verified by measuring the MTF and Wiener spectrum (WS), respectively. The FIR filtering was then applied to the representation of digital bone and chest radiographs.
Results:
The FIR filtering improved the MTF values by up to almost 1.0 or greater over the frequency range of interest, while it minimally increased the WS values. Combining the inverses of the display MTFs with the response of the human eye led to further refinement of the MTF. Our method was successfully and beneficially applied to the image interpretation of bone radiographs. The resolution enhancement of chest radiographs, which include larger scattered radiation than bone radiographs, was easily perceived by incorporating the response of the human eye. In addition, no artifacts were observed on the processed images.
Conclusion:
Our proposed method to compensate for the degradation of the resolution properties of LCDs has the potential to improve the observer performance of radiologists when reading digital radiographs.
This work was supported in part by grant from EIZO Corporation.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.