Measurements obtained with the RTONE, either by an ophthalmologist or by the patient, showed an excellent correlation with those provided by applanation tonometry. RTONE generally tends to overestimate IOP compared to GAT readings and displays a dependence on CCT. This study was registered with the DRKS (German Clinical Trials Register; www.germanctr.de ; DRKS00000478).
Measurements obtained with the RTPRO, either in the upright or in the supine position, show good correlation and agreement with those provided by applanation and dynamic contour tonometry. The study was registered with the DRKS (German Clinical Trials Register; http://www.germanctr.de; DRKS00000581).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.