Background: the efficiency of the diagnostic aids plays an important role in the treatment plan. This study aims to assess the diagnostic accuracy of dental pulse oximeter with a customized sensor holder, thermal test and electric pulp tester in assessing the actual pulp status and to evaluate the oxygen saturation level in control healthy teeth, non-vital and teeth with irreversible pulpitis. Material and methods: thirty-seven single canal teeth requiring endodontic therapy were included in the study. The selected teeth were tested with dental pulse oximeter, electric pulp test, cold spray, and heated gutta percha stick. Between each test a time lag of 2 minutes was allowed for the central sensitization to occur. Three blinded operators were involved in the study. The actual status of the pulp was evaluated after the initiation of endodontic treatment. To assess the actual pulp status direct visual examination of access cavity as performed. The data was statistically analysed using (ANOVA) Analysis of Variance and Post-hoc Tukey test. Results: sensitivity of pulse oximeter, heat test, cold and electric pulp test, was 100, 25, 50, and 12, respectively. The specificity of these tests was 100, 72 81, and 77, respectively. The ANOVA showed that there was statistical difference between all the groups (p=0.0005). Post-Hoc Tukey revealed that there was statistical difference among all the groups, nonvital group (p=0.0005), control group (p=0.01) and for irreversible pulpitis (p=0.01). The overall diagnostic accuracy of pulse oximeter was 100% followed by cold test 66%, heat test to be 49% and electric pulp test to be 45%. Conclusion: the custom-made holder used in the present study aided in providing accurate response for pulp vitality testing. In this study the diagnostic accuracy was high with dental pulse oximeter followed by cold, heat and the least was electric pulp tester in different pulpal conditions.KEYWORDS Cold test; Heat test; Oxygen saturation; Pulse oximeter; Vitality tests.
Aim of the studyTo determine the factors that are likely to influence the domains of health-related quality of life (HRQOL) using SF-36 and CLDQ questionnaires in patients with liver cirrhosis.Material and methodsPatients with liver cirrhosis were compared with age- and gender-matched healthy controls for physical and mental components of the SF-36 score. Effects of age, co-morbidity, namely diabetes, severity of liver disease and complications of liver cirrhosis on HRQOL using self-administered or by direct interview SF-36 and CLDQ questionnaires were studied. Statistical analysis: chi square test, ANOVA, Kruskal-Wallis test and stepwise linear regression analysis. A p value of < 0.05 was considered significant.ResultsRegarding SF-36 score, except for bodily pain, 149 patients had significantly low individual and composite domain scores (p value < 0.0001) compared to age/gender-matched controls. Patients below 45 years, the majority of whom belonged to Child-Turcotte-Pugh (CTP) class C with a high Model of End-Stage Liver Disease (MELD) and higher rates of complication had low SF-36 for bodily pain (KW p < 0.005) and those above 55 years for physical function (p < 0.05). Both the physical components had a major impact on mental composite score (MCS) (KW p < 0.05). The overall CLDQ score was also low in patients below 45 years old (p < 0.05). Diabetes with or without other co-morbid conditions had no effect on SF-36 or CLDQ scores, while non-diabetic co-morbid conditions did on physical domains (physical function, bodily pain and role physical) and the physical component score of SF-36 (KW p < 0.01 to < 0.0001). In linear regression, MELD had a direct and significant association with overall PCS, MCS and CLDQ.ConclusionsAge below 45 years, higher MELD and CTP score with the presence of ascites and hepatic encephalopathy affect the overall CLDQ scores.
Aim:
This study aims to assess the diagnostic accuracy of new custom-made pulse oximeter sensor holder in assessment of actual pulp status with three pulp sensibility tests.
Materials and Methods:
Seventy-nine single canal teeth requiring endodontic therapy was included in the study. The tooth that was requiring root canal treatment was tested with heat test, cold test, electric pulp test, and pulse oximeter. Between each test, a time period of 2 min was allowed. The response from three pulp sensibility tests and the reading from pulse oximeter were recorded. Following which root canal treatment was performed. The result obtained from four pulp tests were correlated with the clinical finding after access cavity preparation. The data obtained was statistically assessed. Receiver operator characteristic (ROC) curve analysis was performed to assess the efficacy of the pulp tests. In the above statistical tools, the probability value. 05 is considered as significant level.
Results:
The overall diagnostic accuracy was found to be significantly higher with pulse oximeter when compared with other three pulp sensibility tests. The ROC curve demonstrates the results obtained from pulse oximeter was found to be more reliable than other pulp tests.
Conclusion:
Within the limitation of the study, diagnostic accuracy of pulse oximeter with custom made sensor holder was reliable and accurate in assessment of actual pulp status.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.