To avoid or minimize postimplantation injury as a result of brain micromotion relative to the skull, a flexible multichannel polyimide (PI) cable was designed and microfabricated for data and power transmission between an intracranial IC recording from a neural probe array and an extracranial IC exchanging power and data wirelessly with an external unit. Surface characteristics, electrical properties, and cytocompatibility of the PI ribbon cable were investigated in this study. Scanning electron microscopic examination and atomic force microscopy analyses showed that the surface of the PI ribbon cable became significantly rougher due to the reactive oxygen ion etching process to open bonding pads. The enhanced surface roughness was also responsible for the increase in wettability and water absorption rate. However, water permeability measurement revealed that the micromachining fabrication process did not meaningfully affect the acceptable water vapor transmission rate of PI. Moreover, electrical properties, such as insertion loss, isolation between channels and data transmission capacity, were assessed for each channel of the PI ribbon cable on the basis of scattering parameter (S-parameter) measurement. Finally, 3-(4, 5-dimethylthiazol-2-yl)-2, 5-diphenyltetrazolium bromide assay and live/dead intracellular staining tests were carried out to evaluate cell behaviors on the PI ribbon cable, indicating that the PI ribbon cable did not have acute cytotoxicity and appeared to be as cytocompatible as blank PI foils.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.