The proposed HDR-BT phantom and instrumented applicator have been tested and their main characteristics have been evaluated. These systems perform unsupervised measurements and analysis without prior treatment plan information. They allow independent verification of dwell position and time with accuracy of measurements comparable with other similar systems reported in the literature.
The intrinsic thermoluminescence (TL) efficiency of a TL detector relates the absorbed dose in the detector material to the light yield observed upon evaluation. Knowledge of the TL efficiency is of interest when performing numerical simulations of detector response, where only absorbed dose can be predicted. Here, the experimental determination of TL efficiency for calcium sulphate (CaSO(4):Tm) and lithium borate (Li(2)B(4)O(7):Cu) is reported. These materials are widely used in Panasonic dosemeter badges. The results of the study are in agreement with predictions from track structure theory and microdosimetry, relating an enhanced light yield at low X-ray energies to supralinear behaviour of the TL phosphor.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations鈥揷itations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.