The Nucleus CI24RE ‘Freedom' device offers higher stimulation rates and lower noise levels in action potential measurements (ECAPs) than previous devices. A study including ten European implant teams showed that the effect of changes in rate from 250 to 3500 pulses per second on tilt and curvature of the T and C profiles is insignificant. When changing rate one may change the levels at all electrodes by the same amount. Using an automated procedure ECAPs could be measured quickly and reliably at a noise level of only 1 microV, this did not result in improved correlations between the tilt and curvature parameters of the ECAP profiles and those of the T and C profiles. Average C levels appear to differ markedly among implant centers; a better assessment protocol is required. When increasing stimulus rate one should take into account that this requires higher pulse charges per second and more power consumption.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.