SummaryA method for the determination of plasma oxalate using oxalate oxidase (EC 1.2.3.4) and deproteinised plasma is described. Reference values are 1.2-6.4 pmol/l (n = 24, mean Lf SD 3.3 + 1.5 pmol/l). The sensitivity is 6-7 nmol, the accuracy 3-5 nmol, and the coefficient of variation 10.4% (at a level of 23 nmol). The recovery from plasma spiked with oxalate was 105 k 8% (n = 8). Allowing blood to stand for 4 h at room temperature had no effect on plasma oxalate levels; inhibitors of glyoxylate converting enzymes interfered with oxalate oxidase.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.