A tubular cation-exchange membrane Is Installed at the end of a 60-cm-long 75-gm-bore fused-slllca capillary. A static dilute acid regenerant solution surrounds the membrane that functions as a suppressor. With positive high voltage applied to the capillary Inlet and the regenerant solution grounded, effective suppression of electrolytes such as solutions of alkali metal borate, glyclnate, or cyanide Is observed. Electroosmotlc flow carries the capillary effluent past the suppressor Into a conductivity detection cell constituted by two platinum wires Inserted through the wall of a poly(vinyl chloride) capillary. The system provides detection limits In the 10-20 pg/L range for a variety of anions; a typical separation requires 15 min. Applicability to a variety of real samples Is demonstrated.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.