A method for the quantification of the natural hormone 17 beta-estradiol (17 beta-E2) in bovine serum by liquid chromatography atmospheric pressure chemical ionization tandem mass spectrometry (LC-APCI-MS-MS) was developed. Ethinylestradiol (EE2) was used as internal standard. Analytes were extracted from serum using acetate buffer, purified by C18 solid-phase extraction (SPE) and chromatographed on a polymeric reversed-phase (PLRP-S) LC column. They were ionized in a heated nebulizer (HN) interface operating in the negative ion mode, where only the intact deprotonated molecules, [M - H]-, were generated at m/z 271 and 295 for 17 beta-E2 and EE2, respectively. These served as precursor ions for collision-induced dissociation (CID) and diagnostic product ions were identified for the unambiguous hormone confirmation by selected reaction monitoring (SRM) LC-APCI-MS-MS. The method was validated on bovine serum and the limit of quantification (LOQ) was 30 pg ml-1 for 17 beta-E2. The inter-day precision (relative standard deviation, RSD) and accuracy (relative error, RE) derived from the analyses of validation samples at three concentrations ranged from 1.76 to 3.76 and from -4.18 to -2.01%, respectively. This method is currently being successfully applied to measure the bovine serum concentration of 17 beta-E2 in order to discriminate between the physiological concentrations of 17 beta/E2 and the hormone levels resulting from illegal administration.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.