A method for the determination of formaldehyde by flow injection analysis with spectrophotometric detection is proposed, based on retarding the reaction between brilliant green and sulphite by the addition of formaldehyde; this was investigated for formaldehyde quantification in extracts from wood-based panels. For the first time, a heating step was explored, providing a sample throughput of 50 analyses per hour, with a limit of detection of 0.02 mg L −1 and linearity of 0.20-3.0 mg L −1 , which was adequate for the expected range of formaldehyde concentration in the extracts. The mean recovery observed for actual samples was in the range of 92-106 %, with a maximum relative standard deviation of 6.0 %. The paired t-test revealed no significant difference between this method and the official Nash method, demonstrating an appropriate accuracy and precision; the method is proposed as a simple, fast and inexpensive alternative for the routine determination of formaldehyde in an aqueous medium.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.