The effect of femoral guide rotation in endoscopic anterior cruciate ligament (ACL) reconstruction is reviewed based on a previous report. The effect of varied offsets of the femoral guide (5.5 and 7.0 mm) are described. This is one of the few isometry studies to evaluate the knee through a practically full range of motion (0 degree-120 degrees). A 7-mm offset guide rotated to the 12:00 position yields the best single fiber and graft excursion patterns (P < 0.05). A 5.5-mm offset guide yields inferior single fiber and graft excursion patterns. Single fiber and graft isometry are similar but not identical in endoscopic ACL reconstruction. Centering the single fiber in the tibial tunnel has little effect on isometry patterns, demonstrating that the more posterior tibial positions needed for endoscopic reconstruction are acceptable from an isometry standpoint.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.