In 1998, the combined Interferometer and Ultra Small Angle Neutron Scattering (USANS) instrument S18 at the 58 MW High Flux Reactor at the Institute Laue-Langevin in Grenoble (France) started operation. The instrument has been upgraded to allow more advanced neutron optics experiments for fundamental, nuclear and condensed matter physics. The new supermirror guide along with the two channel cut silicon perfect crystals, configured as an advanced high resolution Bonse-Hart camera, provides optimal intensity conditions. A large range of wavelengths is accessible by variations of the Bragg angle and by using different reflecting planes of a properly cut silicon monochromator block in combination with different channel-cut analyzer crystals. The fine adjustment analyzer system is achieved by an advanced piezo drive, which has an accuracy of 0.036 seconds of arc. The basic features of the USANS camera, which takes advantage of a new tail suppression method, and the results of various test measurements concerning intensities, long term stability and the accessible wavelength range will be presented. A high resolution position sensitive detector can either be used to analyze beam profiles or to perform neutron radiography experiments.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.