Calibration of patient dose monitoring devices in diagnostic radiology has become increasingly important in the light of new legislation that requires monitoring of patient dose against local and national diagnostic reference levels. An investigation was conducted into the long-term stability of 41 dose-area product (DAP) meters over a period of approximately 5 years, to assess the suitability of an annual calibration regimen. For DAP meters fitted to overcouch X-ray tubes, 77% of calibrations were within 10%, whilst for undercouch tubes only 50% of calibrations were within 10%. These findings suggest that annual calibration may be too infrequent. Suitable calibration frequencies for different clinical workloads are discussed.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.