Steam sterilization processes rely on high heat transfer rates from the steam to the medical devices. To guarantee this, all non-condensable gases (NCGs) must be removed from the steam sterilizer before sterilization. The ratios and distribution of the NCGs in the pre-sterilization phases of both a vacuum and a non-vacuum steam sterilizer are investigated. Two CFD models for simulation of the NCGs in the sterilizers were developed and validated by the measurements of a pressure sensor and eight thermocouples inside the steam sterilizer. Furthermore, a novel method was developed to calculate the real ratio of NCGs based on the standard measurement (EN 13060). The results show that the NCGs ratio defined in the EN 13060 did not accurately represent the real ratios of the NCGs.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.