Background: Due to the rotator cuff retear after being surgically repaired, some strategies have been developed. The authors verified that the possibility of polyetheretherketone (PEEK) vented anchors promoted a better clinical and healing process than PEEK solid anchors.
Methods: A prospective and randomized study was designed with 38 patients treated with PEEK anchors, 18 of whom with vented anchors and 20 with solid ones. Demographic, clinical and radiologic data were collected before and during surgery (time 0) and at 12 months of follow-up.
Results: In the final follow-up (12 months), there was no difference in the visual analogic scale (VAS) scale between groups (1.7 points vs 1.9 points; p = 0.731), neither in the DASH score (34.2 points vs 23.9 points; p = 0.268), nor in absolute Constant score (76.9 points vs 77.3 points; p = 0.910). In MRI, 10 patients had their cuff tear healed in the vented group and 15 in the solid group (p = 0.173).
Conclusion: The new designed vented anchors do not add any advantage when compared to solids ones, at least within the first year after surgery.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.