A procedure for development of a pen-like, multi-electrode potentiometric sensing platform is described. The platform comprises a seven-in-one electrode incorporating all-solid state ion-selective and reference electrodes based on the conductive polymer (poly (3,4-ethylenedioxythiophene) (PEDOT)) as an intermediate layer between the contacts and ionselective membranes. The ion-selective electrodes are based on traditional, ionophore-based membranes, while the reference electrode is based on a polymer membrane doped with the lipophilic salt tetrabutyl ammonium tetrabutyl borate (TBA-TBB). The electrodes, controlled with a multichannel detector system, were used for simultaneous determination of the concentration of Pb 2+ and pH in environmental water samples. The results obtained using pH-selective electrodes were compared with data obtained using a conventional pH meter and the average percent difference was 0.3 %. Furthermore, the sensing system was successfully used for lead speciation analysis in environmental water samples.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.