Inductively coupled plasma atomic emission spectrometry (ICP-AES) and inductively coupled plasma mass spectrometry (ICPMS) methods were applied to check the possibility of determination of hafnium in zirconium at a level lower than 100 ppm. A zirconium matrix of hafnium content lower than 10 ppm was obtained using a worked-out separation method exploiting ion exchange on Diphonix resin. Both methods give results in good agreement with each other as well as with those for certified reference material BCR-098 (Zircaloy-4). They were utilized in determination of Hf in the samples collected during separation of microamounts of hafnium from zirconium by the mentioned ion exchange. These results proved the earlier described method of separation on Diphonix resin to be effective even when the initial concentration of hafnium in zirconium decreases from 2.4% to 0.0082%.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.