In this paper, the design of a new type of heart pump is presented. The uniqueness of this pump concept is that the mechanical energy is transferred to the fluid by the rotation of flat disks without blades. Both theoretical and numerical analysis are used to determine the pump design parameters. The pump design parameters are calculated using the Navier-Stokes equations. The application of computational fluid dynamics is used to define the geometric design. The pump head and flow have to be within strictly defined limits to ensure normal blood circulation. The negative impact of the pump on the blood must be minimal (no stagnation and recirculation zones, shear stress in the acceptable range). It is also important to achieve the smallest possible pump volume. For the pump operating point of ∆p = 65 mmHg, Q = 5.43 L/min, and ω = 6000 rpm, the design parameters are inner radius R1 = 12 mm, outer radius R2 = 15 mm, the distance between disk of h = 1 mm, and the number of disks is n = 6. The shear stress in the rotor is in range 46–108 Pa and the pump residence time is 0.0194 s.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.