We compared the sensor performance of 3 commercially available GM systems. A total of 12 patients with type 1 diabetes were included in a single-centre, open-label study in which the sensor performance of the Abbott FreeStyle libre (Abbott), Dexcom G4 Platinum (Dexcom) and Medtronic MiniMed 640G (Medtronic) systems over 12 hours was compared during mimicked real-life conditions (meals, exercise, hypo-and hyperglycaemia). Sensor performance was determined by fulfilment of ISO 15197:2013 criteria, calculating mean absolute relative difference (MARD), and was also illustrated using Parkes error grid and Bland-Altman plots. Sensor performance during changes in metabolic variables (lactate, betahydroxybutyrate, glucagon, non-esterified-fatty-acids) was determined by Spearman's rank correlation coefficient testing. The systems fulfilled ISO 15197:2013 criteria by 73.2% (Abbott), 56.1% (Dexcom) and 52.0% (Medtronic). The MARDs AE standard deviation in the entire glycaemic range were 13.2% AE 10.9% (Abbott), 16.8% AE 12.3% (Dexcom) and 21.4% AE 17.6% (Medtronic), respectively. All sensors performed less accurately during hypoglycaemia and best during hyperglycaemia. We did not observe an influence of metabolic variables on sensor performance.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.