Two streamlined programming procedures are recommended for use in the clinical setting: (1) interpolating across three measured T-levels and three measured C-levels and (2) interpolating across five measured T- or C-levels and using the interpolated profile for fitting of the alternative profile.
The use of higher stimulation rates may provide benefit to some but not all cochlear implant recipients. It is important to optimize the stimulation rate for an individual to ensure maximal benefit. The absence of any changes in T- and C-levels or in tinnitus suggests that higher stimulation rates are safe for clinical use.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.