Background
The ratio of the antiangiogenic factor, soluble fms-like tyrosine kinase 1 (sFlt-1), to the proangiogenic factor, placental growth factor (PlGF), is associated with increased risk of preeclampsia. Here, we describe an analytical evaluation of the Elecsys sFlt-1 and PlGF assays at the first North American site in which they were clinically implemented.
Methods
The analytical evaluation included short- and long-term imprecision, method comparison, accuracy, linearity, sample stability, limit of quantification verification, and measurement uncertainty. Quality indicators were also evaluated, including turnaround time and repeat test frequency.
Results
Short-term (13-day) and long-term (12-month) imprecision for sFlt-1 and PlGF were <4% CV. Method comparison (n = 40) between Roche cobas e602 and e411 exhibited r > 0.99 and bias <10%. sFlt-1/PlGF ratio rule-out cutoffs (≤33 and ≤38) and rule-in cutoffs (>38, >85, and >110) exhibited negative percent agreement and positive percent agreement of 100%, respectively (n = 40). During the first 12 months, 257 orders were placed, repeat test frequency was 17.5%, mean time between repeat orders was 23 days, and 72.0% of results were reported within 2 h from sample receipt when quality control was run continuously.
Conclusions
We describe analytical performance parameters and quality indicators of the Elecsys sFlt-1 and PlGF assays, which was the first North American clinical laboratory site to implement these assays in support of the institution’s high-risk obstetrical unit.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.