Interlocked intramedullary nailing is the treatment of choice for femoral shaft fractures. However, distal locking is a technically challenging part of the procedure that can result in distal femoral malrotation and high radiation exposure. We have tested a robotic procedure for robotic distal locking based on the computation of a drilling trajectory on two calibrated fluoroscopic images. Twenty distal holes were attempted in ten cadaveric femur specimens. Successful screw hole drilling was achieved at the first attempt in each of the ten specimens (20 drill holes in total). No failures were recorded. The average total number of images needed was 6.5 +/- 3.6. The average computation time was 16.5+/- 16.0 seconds. Robotic distal locking was feasible in this test and can be integrated into a fully robotic intramedullary nailing procedure.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.