Levetiracetam is an antiepileptic drug for the treatment of psychiatric patients. In this study, a selective, straightforward, and rapid online heart‐cutting liquid chromatography method was developed for the therapeutic drug monitoring of levetiracetam. This method allows for the determination of levetiracetam in human plasma without complex sample preparation. The mobile phases consisted of 30 mM aq. orthophosphoric acid solution/methanol (70:30) at a flow rate of 1 mL/min for the first system and 10 mM aq. orthophosphoric acid solution/methanol (55:45) at a flow rate of 1 mL/min for the second system. The first separation was carried out on a GL Sciences Intersil ODS‐3 column (4.6 mm × 150 mm, 3 µm) and the second separation was carried out on a Restek Ultra PFPP column (4.6 mm × 150 mm, 5 µm). The detection was carried out at 205 nm for both systems. The method was validated for selectivity and linearity, which were in the 6–60 µg/mL range. Intra‐ and interassay accuracies were <112.6%, and the intra‐ and interassay precisions were <6.4% for all quality control samples. The lower limit of quantitation was 6 µg/mL. The developed method was successfully applied for therapeutic drug monitoring of plasma samples from patients.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.