Commercially available dip strips
based on peroxidase technology
are widely used to semiquantitatively determine the peroxide concentration
in peroxidizable solvents. However, on the basis of the results of
our study, the method can only qualitatively respond to limited types
of peroxides; it also responds to other oxidants, such as NaClO. Therefore,
test results obtained from dip strips should be used for reference
only. Differential scanning calorimetry is recommended as a more reliable
and direct method to assess decomposition risks.
Novel cages based on acetone complex with dimeric units of calix[4]arene dimethoxycarboxylic acid extend along different directions to form polymeric cages, and the polymers interlock to form layer arrays, that in turn antiparallel pack in a bilayer architecture.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.