A beam-spot camera for measurements of x-ray focal spots of accelerators has been evaluated. The device consists of closely packed 0.25-mm-thick lead and 0.25-mm-thick cardboard strips. It is placed on radiographic film with the lamellae parallel to the beam axis and an exposure made. The images were scanned with a microdensitometer. The results indicate that the broadening of the source intensity profile at half maximum is of the order 1 mm, which permits the use of the beam-spot camera for acceptance testing and quality control. Longer tails in the density profile limit the quantitative information that can be extracted from the images.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.