Some decades ago single photon detection used to be the terrain of photomultiplier tube 9 (PMT), thanks to its characteristics of sensitivity and speed. However, PMT has several 10 disadvantages such as low quantum efficiency, overall dimensions, and cost, making 11 (DCR) has been conducted. Our results show a dark count rate increase with the size of 28 the photodiodes and the temperature (at T=22.5°C, the DCR of a 10 µm-photodiode is 29 2020 count.s-1 while it is 270 count.s-1 at T=-40°C for a overvoltage of 800 mV). A small 30 pixel size is desirable, because the DCR per unit area decreases with the pixel size. We 31 also found that the adjustment of overvoltage is very sensitive and depends on the 32 temperature. The temperature will be adjusted for the subsequent experiments.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.