In this work, polymer gel-MRI dosimetry (using VIPAR gels), radiographic film and a PinPoint ion chamber were used for profile measurements of 6 MV x-ray stereotactic beams of 5 and 10 mm diameter. The VIPAR gel-MRI method exhibited a linear dose response up to 32 Gy. VIPAR gels were found to resolve the penumbra region quite accurately, provided that the in-plane image resolution of the related T2-map is adequate (< or = 0.53 mm). T2-map slice thickness had no significant effect on beam profile data. VIPAR measurements performed with a spatial resolution of 0.13 mm provided penumbra widths (80%-20% distance) of 1.34 and 1.70 mm for the 5 and 10 mm cones respectively. These widths were found to be significantly smaller than those obtained with the film (2.23 mm for the 5 mm cone, 2.45 mm for the 10 mm cone) and PinPoint (2.25 mm for the 5 mm cone, 2.52 mm for the 10 mm cone) methods. Regarding relative depth dose measurements, good correlation between VIPAR gel and PinPoint data was observed. In conclusion, polymer gel-MRI dosimetry can provide relatively accurate profile data for very small beams used in stereotactic radiosurgery since it can overcome, to some extent, the problems related to the finite size of conventional detectors.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.