A rapid and sensitive liquid chromatography tandem mass spectrometry (LC-MS/MS) method has been developed for quantitative determination of Tamsulosin (TMS). The analyte was extracted from human plasma by Liquid-Liquid Extraction (LLE) using Methyl Tertiary Butyl Ether (MTBE) and Dichloromethane (DCM). Tolteridone Tartrate (TLT) was used as the internal standard. A Phenomenex RP-18 (50 x 4.6 mm i.d., 5µ) column provided chromatographic separation of the analyte using a mobile phase containing Acetonitrile: 0.01M ammonium formate (pH9.0) (90:10) at a flow rate of 0.8 ml/min with an elution time as low as 3.0 min which was followed by detection with mass spectrometry. The Multiple Reaction Monitoring (MRM) pair (m/z) 409.3/271.4 for TMS and 326.4/147.1 for TLT. Simple isocratic chromatographic conditions and mass spectrometric detection of the method enables the detection of TMS at less than nanogram levels. The proposed method was found to be linear from 0.20-100.08 ng/ ml. The precision and accuracy values are within 10%. The overall recovery of TMS was 90.92 %.
Purpose: To determine naproxen levels in human plasma using a new liquid chromatography-Mass spectroscopy/Mass spectroscopy (LC-MS/MS) method that involves
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.