The Directive 2013/35/EU establishes standards for workers exposed to static and time varying magnetic fields. These limits are based on ICNIRP guidelines expressed in terms of the electric field induced in the body. The complexity of this measurement led to theoretical models being developed. In this study, the experimental evaluation included varying magnetic field exposures for two classes of MRI workers. The measurements are conducted on four different MRI Systems including one 0.35 T, two 1.5 T, and one 3.0 T. Pocket magnetic dosimeters were used and it was carried out during routine conditions, emergency conditions, and cold-head maintenance/substitution. The acquired data has been processed and the corresponding dB/dt curves have been computed as the first time derivative of the dataset. The weighted peak approach was also implemented for the compliance assessment with regulatory limits. The dB/dt peak values have been compared with the reference level (RL) proposed by ICNIRP. The results show that the RL always exceeds during measurements on the 3.0 T scanner and sometimes on 1.5 T. In light of the foregoing, the diffusion of ultra-high field MRI scanners involves the introduction of behavioral rules that could be more useful than a numerical action level.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.