A precision multi-degree-of-freedom measuring (MDFM) system has been developed and implemented for the simultaneous measurement of straightness, pitch, yaw, and roll errors of the moving axes of a CMM. The system is based on the principles of laser alignment and autocollimator. Its measurement principles and the influence of laser beam drifts on its measurement quality have been investigated and some improvement schemes have been implemented. Through the measurements of actual as well as artificially created geometric errors of the CMM, it has been found that the system’s accuracy of measuring straightness error components is better than 1 μm and its accuracy for angular error measurements is better than 0.5 arcsec.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.