ABSTRACT. The performance of the CO 2 -accepting SO-110 ion source at Oxford Radiocarbon Accelerator Unit has been investigated in detail. The purpose was to clarify the possibilities of accelerator mass spectrometry (AMS) radiocarbon measurements with real-time separation, e.g. GC AMS or HPLC AMS. The construction of a gas test injector based on the continuous-flow technique made it possible to characterize the response of the ion source to continuous and pulsed input of CO 2 gas. The source exhibited remarkably good linearity over a wide range of CO 2 -pulse sizes and fast rise time, but the peak shape varied and memory effects were significant. Appropriate tuning of the gas source proved to be critical.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.