PoisonsInhibitors Interferents a b s t r a c tThe resistance of several models of catalytic, workfunction-based metal-oxide-semiconductor and electrochemical hydrogen sensors to chemical contaminants such as SO 2, H 2 S, NO 2 and hexamethyldisiloxane (HMDS) has been investigated. These sensor platforms are among the most commonly used for the detection of hydrogen. The evaluation protocols were based on the methods recommended in the ISO 26142:2010 standard. Permanent alteration of the sensor response to the target analyte (H 2 ) following exposure to potential poisons at the concentrations specified in ISO 26142 was rarely observed.Although a shift in the baseline response was often observed during exposure to the potential poisons, only in a few cases did this shift persist after removal of the contaminants.Overall, the resistance of the sensors to poisoning was good. However, a change in sensitivity to hydrogen was observed in the electrochemical platform after exposure to NO 2 and for a catalytic sensor during exposure to SO 2 . The siloxane resistance test prescribed in ISO 26142, based on exposure to 10 ppm HMDS, may possibly not properly reflect sensor robustness to siloxanes. Further evaluation of the resistance of sensors to other Si-based contaminants and other exposure profiles (e.g., concentration, exposure times) is needed.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.