A novel fabrication process has been described for the development of a hollow stainless steel microneedle array using femto second laser micromachining. Using this method, a complicated microstructure can be fabricated in a single process step without using masks. The mechanical stability of the fabricated microneedle array was measured for axial and transverse loading. Skin histology was carried out to study the microneedle penetration into the rat skin. Fluid flow through the microneedle array was studied for different inlet pressures. The packaging of the microneedle array, to protect the microneedle bore blockage from dust and other atmospheric contaminations, was also considered. Finally, the microneedle array was tested and studied in vivo for insulin delivery to a diabetic rat. The results obtained were compared with the standard subcutaneous delivery with the same dose rate and were found to be in good agreement.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.