This report describes procedures and results of recent outdoor tests in evaluating a prototype single particle fluorescence analyzer (SPFA), an instrument that is being developed to simultaneously monitor ambient concentrations of both biological and nonbiological aerosols using optical techniques. Recent modifications to the SPFA are also noted and discussed. During recent outdoor tests, aerosols were generated containing four different types of biological material: ova albumin, MS-2 phage, Erwinia herbicola vegetative cells and Bacillus subtilis spores. Analysis of the SPFA results indicates that a probability of detection of at least 87% was achieved for target aerosol concentrations as low as a few (1-5) particles/liter. Moreover, comparison with different reference instruments indicate that the SPFA can achieve absolute quantitative detection efficiencies for individual biological aerosols of at least 70% (assuming appropriate alignment and threshold settings are made). During some of the outdoor tests different nonbiological aerosols were also released that generally did not generate any significant fluorescent signals. Although official test scores have not yet been distributed, we provide here the SPFA results for each of the aerosol releases made during the test. Analysis of the relative fluorescent intensities in the visible and UV spectral bands suggests that these data can be used in the future to provide an indication of the type of biological aerosol being released as well. C
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.