Simple, rapid and precise amperometric methods for quantification of ascorbic acid (AA) are presented. Glassy carbon (GC), carbon paste (CP) and modified carbon paste (MCP) electrodes are used for this purpose. MCP electrodes are of CP with 10 % vanadate (V). All electrodes are inserted in a wall-jet device with an Ag/AgCl reference electrode and a platinum auxiliary electrode. This device is coupled to a flow injection analysis (FIA) set-up. Hydrodynamic and amperometric parameters are optimized unvaryingly. GC electrodes present slopes of 4.75 × 10 5 nA L mol -1 (± 6.4 %) under optimum conditions. CP and MCP electrodes show higher sensitivity, with slopes of 6.37 × 10 5 nA L mol -1 (± 6.6 %) and 7.32 × 10 5 nA L mol -1 (± 4.4 %). Linear responses range (1.0-2.0) × 10 -6 to (0.8-1.0) × 10 -5 mol L -1 . Correlation coefficients were > 0.994 and about 48 samples are analysed per hour. Application of the proposed method to the analysis of soft drinks is presented.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.