In general, marker contrast decreased in function of higher frame rates, which was detrimental for marker detection success. In this work, Visicoils with a thickness of 0.075 cm were showing best results for a 15 Hz frame rate, while non-MV compatible 0.05 cm Visicoil markers were not visible on the new EPID with improved sensitivity compared to EPID models based on a Kodak Lanex Fast scintillator. No noticeable influence of pulsing artifacts on the detection of a 0.075 cm Visicoil was observed, while a synchronized readout provided most reliable detection of a marker-less soft-tissue structure.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.