The impact of calcium chloride (CaCl2) on the performance of N-(hydroxymethyl)acrylamide (NHMA) polymer gel dosimeter is studied in this article. The dosimeter was exposed to doses of up to 10 Gy with radiation beam-energy of 10 MV and dose-rates of 300 cGy/min. The relaxation rate (R2) parameter was utilized to explore the performance of irradiated NHMAGAT gels. The dose response in terms of R2 increased from 0.29 to 0.63 Gy−1·s−1 with increasing calcium chloride concentration from 0 to 1000 mM. The results show no substantial impact of dose-rates as well as radiation energies on NHMAGAT samples. For the steadiness of irradiated NHMAGAT dosimeters, it was found that there is no apparent variation in R2 (less than ±3%; standard deviation) up to 3 days. The overall uncertainty of the gel dosimeter with calcium chloride is 4.96% (double standard deviation, 95% confidence level).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.