Qiagen's Investigator™ Quantiplex kit, a total human DNA quantitation kit, has a 200-base pair internal control, fast cycling time, and scorpion molecules containing a covalently linked primer, probe, fluorophore, and quencher. The Investigator™ Quantiplex kit was evaluated to investigate a value under which complete short tandem repeat (STR) failure was consistently obtained. Buccal swabs were extracted using the Qiagen QIAamp(®) DNA Blood Mini Kit, quantified with the Investigator™ Quantiplex kit using a tested half-volume reaction, amplified with the ABI AmpFlSTR(®) Identifiler kit, separated on the 3100Avant Genetic Analyzer, and data analyzed with GeneMapper(®) ID v.3.2. While undetected samples were unlikely to produce sufficient data for statistical calculations or CODIS upload (2.00 alleles and 0.82 complete loci on average), data may be useful for exclusionary purposes. Thus, the Investigator™ Quantiplex kit may be useful for predicting STR success. These findings are comparable with previously reported data from the Quantifiler™ Human kit.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.