In this case report, we describe markedly different pacing thresholds determined by a manual threshold test and the automatic Ventricular Capture Management algorithm. The discrepancy in pacing threshold values reported was due to the difference in the AV intervals used with the different testing methods. We propose that the differences in right ventricular dimensions with altered diastolic filling periods affected the threshold in this patient with a new passive fixation lead in the right ventricular apex.
During biventricular ICD implantation, pacing the epicardial inferior posterior-lateral surface of the left ventricle via the coronary sinus produced asystole. The proposed mechanism is a Bezold-Jarisch reflex manifested by a drop in heart rate and blood pressure through direct C fiber stimulation.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.